product description
Technical Specifications
|
PARAMETER |
SPECIFICATION |
|
Product name |
Low temperature plasma hydrogen peroxide sterilizer |
|
Product form |
Automatic door vertical |
|
Heating method |
Electric heating |
|
Power Volume Chamber size |
VPS-DZ100: 3.5KVA / 100L (Liner size: 700 × 430 × 360mm) VPS-DZ130: 4.2KVA / 135L (Liner size: 750 × 450 × 400mm) VPS-DZ220: 4.8KVA / 190L (Liner size: 820 × 510× 460mm) |
|
Power |
AC220V, 50Hz |
|
Working temperature |
40-60 ℃ |
|
Select sterilizer |
58%-60% hydrogen peroxide (H202) |
|
Sterilization time |
35-55 minutes |
|
Installation environment |
10-40℃ |
|
Temperature display |
LCD touch screen |
|
Maximum anti-stress value |
10pa |
|
Operation height |
≤160cm |
|
Material of Sterilization Room |
Using high-quality 5052 aluminum material, the cavity structure is rectangular, to improve space utilization, cavity thickness ≥ 8mm, has excellent thermal conductivity. |
|
Outer box material |
ABS plastic disposable plastic. |
|
Electrode mesh material |
aluminum alloy material 5052, sheet metal forming, thickness ≥ 2mm. |
|
Sealing door material |
5052 aluminum alloy material, thickness ≥15mm. |
|
Loading method |
The upper and lower two layers of aluminum alloy punching plates are loaded with sterilized items, with large loading capacity and high space utilization rate. |
|
Door opening method |
Adopt a jack-driven electric lifting door and a light-sensing infrared sensor switch to control the opening and closing of the sterilization chamber door. |
|
Door anti-pinch function |
has the function of door barrier switch. |
|
Cavity temperature heating power |
Preheating heating time ≤30min. |
|
Main insulation |
≥20mm rubber and plastic sponge. |
|
Full automatic control system |
LCD circuit touch screen display PLC circuit design, safe and reliable, the system can be maintained and upgraded. |
|
Printer |
A micro-printer is used to print out the data every minute during the entire sterilization process, which can be permanently saved. |
|
PLC |
Use Siemens smart series PLC control system. |
|
Display screen |
A 7-inch color touch screen is used, and the communication rate is ≥ 19.2Kbps. |
|
Display status |
Temperature, pressure, time, cycle mode, process stage and alarm information, etc., and provide actual interface photos. |
|
Storage function |
It can store and record the sterilization data of 500 times. |
|
Biological indicator incubator |
A microcomputer-controlled biological indicator dedicated constant temperature incubator is used for the constant temperature cultivation process of the biological indicator for sterilization effect monitoring. The temperature control accuracy is ≤ ± 0.3 ℃ within the full range; the temperature can be set by itself . |
|
Pressure Sensor |
The built-in pressure sensor adopts imported sensors to automatically detect the pressure in the cavity to ensure that the accuracy of the pressure in the cavity is controlled within the effective range. |
|
Temperature Sensor |
The built-in temperature sensor automatically detects the temperature in the cavity to ensure that the temperature in the cavity is kept within the precise range of 50 ℃ ± 5 ℃. |
|
Program system |
There are two programs to strengthen circulation and rapid circulation. |
|
Network Connection |
There is a network connection port. |
|
Vacuum system |
The vacuum pumps of well-known domestic manufacturers are used, which are simple and convenient, with a fast vacuum rate and a long service life. |
|
Evacuation Control Valve |
Adopt high vacuum baffle solenoid valve to control the evacuation pipeline. |
|
Oil mist filter |
The product has an exhaust oil mist filter system. |
|
Air filter |
The filtration accuracy is less than or equal to 0.22μm. |
|
Plasma power supply |
Adopt transistor control power supply, power ≤800W, with the advantages of stable power output and high frequency accuracy. |
|
Hydrogen peroxide filling method |
Cartridge filling method. |
|
Sterilization effect |
The silicone hose with an inner diameter of 1mm and a length of 4m can achieve the sterilization effect. |
|
Heating System |
Adopt explosion-proof flame retardant heating system. |
|
Various protection functions |
With over-voltage and over-temperature protection functions. |