product description
Feature
-Machine shell: 304 stainless steel , the surface is easy to clean
-Sterilization chamber: Horizontal round chamber made of 304 stainless steel, covered
with external thermal insulation layer.
-Door: Mechanical and electronic double lock, electronic safety interlocks prevent the
opening of the door if the chamber under pressure.
-Trays : two removable grill in a basket for items load
-Built-in electric steam generator. Equipment with water level sensor, when water lack will have a water lack alarm on screen.
-Safety protection against over pressure, over temperature.
-Vacuum pump: water-cooled oil-free vacuum pump, fast vacuum speed, free maintenance.
-Pipe lines: stainless steel pipe lines for water and steam.
-With Printer
Technical Specifications
PARAMETER |
SPECIFICATION |
Product form |
Vertical |
Heating method |
Electric heating |
Power |
18.0KVA |
Power |
AC380V, 50Hz |
Dimensions |
(L × W × H): 1360x810x1580mm |
Liner size |
(Diameter × Depth): 500 × 1000mm |
Capacity |
200L |
Shelves |
Two trays |
Weight |
360kg |
Temperature display |
7” LCD Touch screen |
Design Pressure |
0.3MPa |
Working pressure |
-0.8MPa~0.21MPa |
Design temperature |
150 ℃ |
Working temperature |
121 ℃ / 134 ℃ |
Number of pulses |
3 times (0~400 times adjustable) |
Sterilization time |
0 ~ 9999 seconds |
Drying time |
0 ~ 9999 seconds |
Safety valve in sterilization chamber |
Open ≧ 0.25Mpa, close ≦ 0.24Mpa |
Pulsation amplitude |
-80.0KPA~150KPA |
Water source of evaporator |
softened or purified water, water temperature ≤45 ℃, water pressure 0 ~ 0.05 Mpa |
Water source of vacuum pump |
Tap water (hardness should not be too high), water pressure 0.15Mpa ~ 0.30Mpa |
Air Discharge |
> 99 ﹪ (Pulsation times more than 3 times) |
Vacuum Drying |
Avoid humidifying the disinfection package after disinfection |
Pulse exhaust |
Positive pressure dynamic pulse exhaust to completely eliminate cold air in the sterilization room and ensure steam saturation |
Working procedures |
Water injection, preheating, pulse vacuum, sterilization, drying and dormancy are automatically completed |
Microcomputer control technology |
The sterilizer uses microprocessor intelligent control, man-machine interface, easy to operate, safe and reliable. The parameters and status during operation are displayed dynamically through the digital screen. The equipment works abnormally and automatically detects and alarms after sterilization is completed, the system automatically prompts and is accompanied by a buzzer warning sound |
Locking device |
Before the start of the sterilization cycle, the interlocking device is not locked after the door is closed, and the door can be closed and opened repeatedly. Only when the door is closed in place, the sterilizer can start to run. During the operation, the sterilizer door should not be opened. After the start of the cycle operation, if a fault or fault indication occurs, the sterilization chamber is not completely released, and when the interlock device should be locked, the sterilizer door should not be opened |
Pulsation mode |
Microcomputer control automatically selects the number of pulsations according to the site conditions to ensure the coordination of temperature and pressure |
Vacuum mode |
Three pre-vacuums to ensure faster heating and more complete sterilization |
Safety valve |
When the pressure of the sterilizer reaches 240kPa ± 5kPa, the safety valve will automatically open to relieve the pressure |
Over pressure protection program |
When the pressure in the sterilization chamber reaches 220 ± 5kPakPa, there is a prompt sound and the pressure can be automatically released |
Air Leakage |
When the sterilizer is tested in various stages, the pressure increase rate should not be greater than 0.13kPa / min |
Dynamic pressure in sterilization chamber |
The pressure change in any 2s interval during the sterilizer cycle should not exceed 1000kPa / min |
Vacuum pump |
The water ring vacuum pump is used, which is simple and convenient, the vacuum rate is fast, and the service life is long. Ensure that the vacuum can reach -80kpa |
Pressure Sensor |
The built-in pressure sensor automatically detects the pressure in the cavity |
Temperature Sensor |
The built-in temperature sensor automatically detects the temperature in the cavity to ensure that the cavity is kept within an accurate range of ± 3 ℃ of the operating temperature |
Sterilization Mode |
Sterilization mode for solid equipment, dressing packaging, rubber products, BD test, high pressure, etc |
Automatic drying |
Powerful vacuum drying, suitable for different items drying needs; the remaining humidity of the equipment is <0.2% |
Seal structure |
The sealing ring is made of silicone, and the designed service life is 2 years (1200 times) without aging and deformation |
Foot wheel |
Yes |
Alarm system |
High and low temperature alarm system |